Title of article
The recoil proton telescope in non-coincidence mode for neutron fluence measurements
Author/Authors
Lِvestam، نويسنده , , Gِran، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
6
From page
609
To page
614
Abstract
The paper describes an alternative approach to use the Los Alamos-type Recoil Proton Telescope for fluence measurements of mono-energetic neutrons with energies above 1 MeV. Traditionally, this type of detectors are used in a 3-coincidence acquisition mode (proportional counter–proportional counter–end detector). This study shows that by only using the end detector and a background reduction method similar results are obtained, with considerably less complexity in the measurement set-up but with larger uncertainties.
Keywords
Recoil proton telescope , Neutron fluence detector