Title of article
Proton damage in the E2V swept charge device
Author/Authors
Holland، نويسنده , , A.D. and Hutchinson، نويسنده , , I.B and Smith، نويسنده , , D.R and Pool، نويسنده , , P، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
6
From page
393
To page
398
Abstract
The swept charge device has been developed using CCD technology to perform photon counting X-ray spectroscopy in the 0.5–10 keV band as an alternative to silicon PIN and drift diode detectors. The key benefits of the detector are large area, combined with reduced electrical interconnects, with X-ray performance obtained using modest cooling to around −20°C. Here we test the technology for its suitability for space use where solar protons degrade the CTI of the detector leading to a degradation in its resolution as a spectrometer. We report on the irradiation experiments conducted and discuss changes in leakage current and energy resolution in the detector.
Keywords
Charge coupled device , CCD , Proton damage , X-ray spectrometer , Radiation damage
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2004
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2202131
Link To Document