Title of article :
A simulation of portable PET with a new geometric image reconstruction method
Author/Authors :
Kawatsu، نويسنده , , Shoji and Ushiroya، نويسنده , , Noboru، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
3
From page :
445
To page :
447
Abstract :
A new method is proposed for three-dimensional positron emission tomography image reconstruction. The method uses the elementary geometric property of line of response whereby two lines of response, which originate from radioactive isotopes in the same position, lie within a few millimeters distance of each other. The method differs from the filtered back projection method and the iterative reconstruction method. The method is applied to a simulation of portable positron emission tomography.
Keywords :
PET , Image reconstruction
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2006
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2202302
Link To Document :
بازگشت