• Title of article

    Automated wire tension measurement system for LHCb muon chambers

  • Author/Authors

    Ciambrone، نويسنده , , P. and Dané، نويسنده , , E. and Dumps، نويسنده , , R. and Dwuznik، نويسنده , , M. and Felici، نويسنده , , G. and Forti، نويسنده , , C. and Frenkel، نويسنده , , A. and Graulich، نويسنده , , J.-S. and Kachtchouk، نويسنده , , A. and Kulikov، نويسنده , , V.V. and Martellotti، نويسنده , , G. and Medvedkov، نويسنده , , A. and Nedosekin، نويسنده , , A.A. and Penso، نويسنده , , G. and Pinci، نويسنده , , D. and Pirozz، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    8
  • From page
    156
  • To page
    163
  • Abstract
    A wire tension meter has been developed for the multi-wire proportional chambers of the LHCb muon detector. The wire tension is deduced from its mechanical resonance frequency. In the LHCb muon chambers, the wires are 2 mm apart and electrically connected in groups of 3–32, so that the wire excitation system must be precisely positioned with respect to the wire to be tested. This wire is forced to oscillate by a periodic high voltage applied between that wire and a non-oscillating “sense wire” placed parallel and close to it. This oscillation produces a variation of the capacitance between these two wires which is measured by a high precision digital electronic circuit. At the resonance frequency this capacitance variation is maximum. The system has been systematically investigated and its parameters were optimized. In the range 0.4–1 N a good agreement is found between the mechanical tension measured by this system and by a dynamometer.
  • Keywords
    Multi-wire proportional chambers , Elementary-particle and nuclear physics experimental methods and instrumentation , Tracking and position-sensitive detectors
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2005
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number