• Title of article

    Quality of X-ray transmission radiography based on single photon counting pixel device

  • Author/Authors

    Jakubek، نويسنده , , Jan and Vavrik، نويسنده , , Daniel and Pospisil، نويسنده , , Stanislav and Uher، نويسنده , , Josef، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    5
  • From page
    113
  • To page
    117
  • Abstract
    This work investigates the optimization of digital X-ray transmission radiography for the observation of time-dependent processes using a single X-ray photon counting pixel device. The maximization of frame rate of acquired roentgenograms is crucial for this purpose. The influence of the X-ray tube voltage on the exposure time needed to acquire images of sufficient quality was investigated. Image distortions due to non-uniform pixel detection efficiency across the X-ray detector can be successfully eliminated by a “Direct thickness calibration method” proposed here, in spite of X-ray beam hardening due to the variation of inspected body absorption.
  • Keywords
    digital radiography , Roentgenogram quality , Single photon counting pixel device
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2005
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2203689