Title of article :
Origins of the residual pulse height deficit in propane-filled gas ionization detectors
Author/Authors :
Weijers-Dall، نويسنده , , T.D.M. and Timmers، نويسنده , , H. and Elliman، نويسنده , , R.G.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
6
From page :
139
To page :
144
Abstract :
This work investigates the origins of the residual pulse height deficit in gas ionization detectors. It is motivated by the recent observation that the species dependence of gas detector response cannot be accounted for solely by considering the energy loss of the ions in the detector window and non-ionizing energy loss processes in the detector gas. It was found that the residual pulse height deficit is approximately proportional to the square of the ionization density. However, only a weak dependence of the residual deficit on gas pressure (in the range 70– 120 mbar ) was observed. It is hypothesized that the residual pulse height deficit in gas ionization detectors results from the effect of multiple ionization of individual gas molecules at high ionization densities on the energy required to create an electron–ion pair.
Keywords :
Elastic recoil detection , Detector response , Gas ionization detector , Ion beam analysis , Pulse height deficit
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2005
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2204001
Link To Document :
بازگشت