Title of article :
Quality evaluation of CsI photocathodes for the ALICE/HMPID detector
Author/Authors :
Hoedlmoser، نويسنده , , H. and Braem، نويسنده , , A. and De Cataldo، نويسنده , , G. and Davenport، نويسنده , , M. and Di Mauro، نويسنده , , A. and Franco، نويسنده , , Jason A. C. Gallas، نويسنده , , A. and Martinengo، نويسنده , , P. and Nappi، نويسنده , , E. and Piuz، نويسنده , , F. and Schyns، نويسنده , , E.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
6
From page :
140
To page :
145
Abstract :
The facility for the production of the large area ( 64 × 42 cm 2 ) CsI photocathodes for the ALICE/HMPID detector has been equipped with a photo-current scanner system for the in situ measurement of the CsI response over the full photosensitive area before transfer to the detector. The photo-current measured on a first batch of 17 PCs, out of the 42 needed to fully equip the seven HMPID modules, has been correlated with test beam measurements in order to define acceptance criteria. Furthermore the systemʹs sensitivity to variations in quantum efficiency allows us to study various effects such as post-deposition heat enhancement or decreased quality due to ageing.
Keywords :
rich , HMPID , CsI photocathode , Heat enhancement , ageing
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2005
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2204228
Link To Document :
بازگشت