Title of article :
Si-APD readout for LaBr3:Ce scintillator
Author/Authors :
Scafè، نويسنده , , Raffaele and Pani، نويسنده , , Roberto and Pellegrini، نويسنده , , Rosanna and Iurlaro، نويسنده , , Giorgia and Montani، نويسنده , , Livia and Nerina Cinti، نويسنده , , Maria، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
3
From page :
355
To page :
357
Abstract :
In this work, we report and compare the results of γ-ray spectroscopy measurements performed using a LaBr3:Ce crystal coupled to a photomultiplier tube with modified base and to a Si-avalanche photodiode readout by standard electronics. Energy resolution values have been obtained in the range from 59.5 keV (241Am) to 1332 keV (60Co). Good linearity in energy response has been found. Energy resolution values obtained with avalanche photodiode readout are underestimated due to the small active area of the presently available photosensor with respect to crystal dimension.
Keywords :
Nuclear medicine , ?-Ray spectroscopy , scintillator , LaBr3:Ce