Title of article :
An instrument for low-level measurements of the leakage current from high-voltage biased detectors
Author/Authors :
Arnaboldi، نويسنده , , C. and Bertoni، نويسنده , , R. and Delucia، نويسنده , , A. C. Pessina، نويسنده , , G. and Redaelli، نويسنده , , N. and Tabarelli، نويسنده , , T.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
2
From page :
193
To page :
194
Abstract :
Resistive Plates Chambers (RPC) are detectors biased at High-Voltage (HV) in excess of 4 kV. When fired by a particle, they develop a large signal current that can be read across a small resistance, 100 Ω or so. A characterization has been made of their ageing as a function of the behaviour of their leakage current with time. An array of 10 detectors has been developed for this purpose. We present the instrument designed and built to perform a continuous and automatic monitoring of the leakage current from each detector of the array, while the system is taking data. For the particular biasing set-up adopted, the current has been measured in series to the terminal connected to the HV of every channel. Since the small value of the currents, order of tens of nA, a special circuit solution and special precautions have been adopted.
Keywords :
Low-level current measurements , Resistive Plate Chambers (RPC) , Detectors , High-voltage monitoring
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2007
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2204997