Abstract
We extend our technique for observing very small concentrations of impurities in xenon gas to the problem of krypton detection. We use a conventional mass spectrometer to identify the krypton content of the xenon, but we improve the sensitivity of the device by more than five orders of magnitude with a liquid nitrogen cold trap. We find that the absolute krypton concentration in the xenon can be inferred from the mass spectrometry measurements, and we identify krypton signals at concentrations as low as 0.5×10−12 mol/mol (Kr/Xe). This technique simplifies the monitoring of krypton backgrounds for WIMP dark matter searches in liquid xenon.
Keywords
krypton , Xenon , Cold trap , mass spectrometry , Dark matter