Title of article :
Pulse shape discrimination in hydrogen-filled proportional counters by digital methods
Author/Authors :
Hawkes، نويسنده , , N.P.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
4
From page :
133
To page :
136
Abstract :
A fast digital sampler has been employed in place of conventional analogue electronics to measure the rise times and amplitudes of pulses from a hydrogen-filled spherical proportional counter of the SP2 type. The rise-time information was used to distinguish neutron events from gammas, and separated neutron events were recorded at energies below those that would have been possible with amplitude measurements alone. The electronics were considerably simpler than those required for an equivalent analogue system.
Keywords :
Neutron spectrometry , Hydrogen-filled proportional counters , Pulse shape discrimination , Digital sampling
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2007
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2205661
Link To Document :
بازگشت