Title of article :
Development of a soft X-ray pulse height analyzer on the J-TEXT tokamak
Author/Authors :
Jin، نويسنده , , W. and Zhuang، نويسنده , , G. and Ding، نويسنده , , Y.H. and Zhang، نويسنده , , Y. and Zhang، نويسنده , , J.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Pages :
5
From page :
15
To page :
19
Abstract :
A soft X-ray pulse height analyzer has been designed and installed on the J-TEXT tokamak for measuring the electron temperature and possibly for monitoring the impurity content and concentration. A multi-channel analyzer, a set of amplifiers, and three silicon drift detectors constitute the detection system. An off-line analysis code was developed to analyze the raw data recorded by the system. The system can measure a minimum electron temperature of 400 eV, and its routine operation provides electron temperature measurements at three different positions with a temporal resolution of 60 ms and channel separation of 18 cm. Line radiation from some moderate and high Z impurities are observable by the system. The results show that the intensity of line radiation from heavy impurity ions is inversely proportional to the plasma density under similar discharge conditions.
Keywords :
pulse height analysis , Line radiation , Soft X-ray , tokamak