Title of article
Minimization of the measurement errors induced by the cyclotron magnetic field measurement system
Author/Authors
?irkovi?، نويسنده , , Sa?a and Ili?، نويسنده , , Andjelija ?. and Dobrosavljevi?، نويسنده , , Aleksandar and Balvanovi?، نويسنده , , Roman and Risti?-Djurovi?، نويسنده , , Jasna L.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2012
Pages
7
From page
54
To page
60
Abstract
Systematic errors caused by the deficiencies of the measurement equipment are occurring relatively often in the engineering practice. The magnetic field measurement system of the VINCY Cyclotron has been designed well; however, due to the practical limitations in the machining process fabrication errors are imminent. We present our experiences with the produced measurement system and the techniques used for the detection and correction of errors. A change in the measuring protocol is suggested in order to overcome otherwise unavoidable errors caused by measuring system machining imperfections.
Keywords
Systematic error , Cyclotrons , Ion beams , Data analysis , Magnetic field measurement
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2012
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2205998
Link To Document