Title of article :
Role of electrode metallization in the performance of bulk semi-insulating InP radiation detectors
Author/Authors :
Za?ko، نويسنده , , B. and Dubeck?، نويسنده , , F. and Proch?zkov?، نويسنده , , O. and Ne?as، نويسنده , , V.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
5
From page :
98
To page :
102
Abstract :
This work deals with the study of three different electrode metallizations with the aim to form a Schottky barrier contact. Electrode geometry corresponds to the requirements of digital radiography systems. As substrates bulk Liquid Encapsulated Czochralski (LEC) SI InP wafers doped with Fe and Fe+Zn are used. Results of this study show that no one of the used metallization performs as a blocking contact. However, detectors with Ti/Pt/Au metallization attained a relatively good energy resolution of 7.0 keV in full-width at half-maximum (FWHM) and the charge collection efficiency (CCE) higher than 83% for 122 keV γ-photons at 255 K. The development of SI InP radiation detectors and in particular their electrode technology is discussed in the light of observed results.
Keywords :
semi-insulating , radiation detector , InP , ?-detection
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2007
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2206077
Link To Document :
بازگشت