Author/Authors :
Gallin-Martel، نويسنده , , M.-L. and Grondin، نويسنده , , Y. and Gac، نويسنده , , N. and Carcagno، نويسنده , , Y. and Gallin-Martel، نويسنده , , L. and Grondin، نويسنده , , D. and Marton، نويسنده , , M. and Muraz، نويسنده , , J.-F and Rossetto، نويسنده , , O. and Vezzu، نويسنده , , F.، نويسنده ,
Abstract :
A detector with a very specific design using liquid Xenon (LXe) in the scintillation mode is studied for Positron Emission Tomography (PET) of small animals. Two prototype modules equipped with Position Sensitive Photo Multiplier Tubes (PSPMTs) operating in the VUV range (178 nm) and at 165 K were built and studied in coincidence. This paper reports on energy, time and spatial resolution capabilities of this experimental test bench. Furthermore, these experimental results were used to perform the first image reconstruction of a 22Na source placed in the experimental setup.