Title of article :
An improved method to accurately calibrate the gantry angle indicators of the radiotherapy linear accelerators
Author/Authors :
Chang، نويسنده , , Liyun and Ho، نويسنده , , Sheng-Yow and Du، نويسنده , , Yi-Chun and Lin، نويسنده , , Chih-Ming and Chen، نويسنده , , Tainsong، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
5
From page :
441
To page :
445
Abstract :
The calibration of the gantry angle indicator is an important and basic quality assurance (QA) item for the radiotherapy linear accelerator. In this study, we propose a new and practical method, which uses only the digital level, V-film, and general solid phantoms. By taking the star shot only, we can accurately calculate the true gantry angle according to the geometry of the film setup. The results on our machine showed that the gantry angle was shifted by −0.11° compared with the digital indicator, and the standard deviation was within 0.05°. This method can also be used for the simulator. In conclusion, this proposed method could be adopted as an annual QA item for mechanical QA of the accelerator.
Keywords :
Calibration of gantry angle indicator , Mechanical quality assurance
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2007
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2206190
Link To Document :
بازگشت