• Title of article

    Scatter radiation influences in nuclide calibration fields

  • Author/Authors

    Hranitzky، نويسنده , , C. and Stadtmann، نويسنده , , H.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    4
  • From page
    230
  • To page
    233
  • Abstract
    The goal of this study is to characterize the 137Cs and 60Co nuclide calibration fields of the reference irradiation facility with collimated beam geometry as part of the Dosimetry Laboratory Seibersdorf. Air kerma along the main beam axis with source–detector distances up to 19 m and additionally lateral field profiles are investigated by comparing mainly ionization chamber measurements with MCNP Monte Carlo computer simulations. The scatter radiation contributions from various facility components including simulations with and without air environment are studied by detailed modeling of the facility. Relative differences between measured and simulated results are currently within about 1–2%, i.e., roughly two to three times of the total measurement uncertainties, but reasons due to scattering influences and further improvements in the model are discussed.
  • Keywords
    Monte Carlo , MCNP , SIMULATION , Calibration , Scatter radiation
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2007
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number