Title of article
Spectrometric measurements of 131I and 99mTc activity in thyroid
Author/Authors
O?ko، نويسنده , , J. and Golnik، نويسنده , , N. and Pliszczy?ski، نويسنده , , T.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
4
From page
578
To page
581
Abstract
A modified method for calibration of the thyroid counter with 131I was investigated in order to take into account the depth of the thyroid in human neck. The depth of the 131I source was determined from the ratio of the counting rate in 364 keV peak to the counting rate in the energy band of 110–140 keV. Dependence of detector efficiency on the source depth was determined. The results were used for analysis of the energy spectra recorded for hospital patients. Similar method was investigated also for 99mTc radiation source.
Keywords
Activity measurements , Thyroid , iodine , Technetium
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2007
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2207597