Author/Authors :
Braem، نويسنده , , A. and Chesi، نويسنده , , E. and Joram، نويسنده , , C. and Seguinot، نويسنده , , J. and Weilhammer، نويسنده , , P. and De Leo، نويسنده , , R. and Nappi، نويسنده , , E. and Lustermann، نويسنده , , W. and Schinzel، نويسنده , , D. and Johnson، نويسنده , , I. and Renker، نويسنده , , D. and Albrecht، نويسنده , , S.، نويسنده ,
Abstract :
We describe a novel method to extract the axial coordinate from a matrix of long axially oriented crystals, which is based on wavelength shifting (WLS) plastic strips. The method allows building compact 3-D axial gamma detector modules for PET scanners with excellent 3-D spatial, timing and energy resolution while keeping the number of readout channels reasonably low. A voxel resolution of about 10 mm3 is expected. We assess the performance of the method in two independent ways, using classical PMTs and G-APDs to read out the LYSO (LSO) scintillation crystals and the WLS strips. We observe yields in excess of 35 photoelectrons from the strips for a 511 keV gamma and reconstruct the axial coordinate with a precision of about 2.5 mm (FWHM).
Keywords :
PET , Santillator , HPD , G-APD