• Title of article

    Transparent target for ultrafast electron beam tomography

  • Author/Authors

    Hampel، نويسنده , , U. and Barthel، نويسنده , , F. and Bieberle، نويسنده , , M. and Stürzel، نويسنده , , T.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    5
  • From page
    8
  • To page
    12
  • Abstract
    In electron beam tomography a rotating X-ray source is generated by electromagnetic scanning of a high-power electron beam across a spatially fixed circular metal target. The X-rays passing the object of investigation are registered by a circular fast X-ray detector array. A basic problem associated with this scanning principle is the angular overlap of the X-ray focal spot path and the detector, which is required for image reconstruction. Therefore, the X-ray production target and the detector array are commonly arranged with some axial displacement. This displacement may produce image artifacts because X-rays are not longer confined to a transversal plane. We propose a novel transparent X-ray target, which fully solves this problem. The new target concept has been practically tested with respect to thermal behavior and electron beam resistance. Its successful application in an experimental tomography setup is demonstrated.
  • Keywords
    electron beam tomography , X-ray target
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2011
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2208114