Title of article :
Optical attenuation measurements in metal-loaded liquid scintillators with a long-pathlength photometer
Author/Authors :
Goett، نويسنده , , Johnny and Napolitano، نويسنده , , James K. Yeh، نويسنده , , Minfang and Hahn، نويسنده , , Richard and Cumming، نويسنده , , J.B. and Hans، نويسنده , , Sunej and Rosero، نويسنده , , Richard and Diwan، نويسنده , , Milind and Sexton، نويسنده , , Ken، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Pages :
6
From page :
47
To page :
52
Abstract :
We describe precise optical attenuation measurements of LAB based liquid scintillators using a custom-built photometer with variable pathlengths up to 2 m. Results are presented for linear alkyl benzene (LAB) and LAB based liquid scintillators at wave lengths of 405, 430, 450, and 470 nm. For these measurements much care was exercised to control the purity of the liquids and additives, as well as the materials of the photometer to control contaminations. These data should be valuable for design of large detectors, such as underground neutrino detectors, using these materials as the media.
Keywords :
Organic scintillator , Neutrino detection , Optical attenuation , Linear alkyl benzene
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2011
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2208166
Link To Document :
بازگشت