• Title of article

    Design, construction, and performance of a magnetically shielded room for a neutron spin echo spectrometer

  • Author/Authors

    Soltner، نويسنده , , Helmut and Pabst، نويسنده , , Ulrich and Butzek، نويسنده , , Michael and Ohl، نويسنده , , Michael and Kozielewski، نويسنده , , Tadeusz and Monkenbusch، نويسنده , , Michael and Sokol، نويسنده , , Don and Maltin، نويسنده , , Larry and Lindgren، نويسنده , , Eric and Koch، نويسنده , , Stuart and Fugate، نويسنده , , David، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    8
  • From page
    40
  • To page
    47
  • Abstract
    A double-layer magnetically shielded room (MSR) has been designed and constructed for the neutron spin echo (NSE) spectrometer at the Spallation Neutron Source (SNS) in Oak Ridge, Tennessee. The primary objective of the MSR is to ensure an undisturbed operation of the spectrometer in terms of external magnetic fields from high-field magnets at neighboring beamlines and from other external devices. Because of the required mobility of the spectrometer along its beamline the MSR features a total length of about 17 m, which makes it the largest MSR worldwide. Several physics and engineering aspects addressed in the design phase and during the construction of this unique MSR are described in this article.
  • Keywords
    Magnetic analysis , Magnetic field effects , Magnetometer , Magnetic shielding
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2011
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number