Title of article :
Improvement of a second-order focusing toroidal spectrometer by use of a pre-collimating lens
Author/Authors :
Hoang، نويسنده , , H.Q. and Khursheed، نويسنده , , A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Pages :
3
From page :
245
To page :
247
Abstract :
This paper presents a method for improving the performance of a second-order focusing toroidal spectrometer using a pre-collimating lens. Simulation results predict that the degree to which the energy resolution can be improved depends on the relative size of the lens. For a pre-collimating lens having a radius 180 times smaller than the spectrometer radius, the relative energy resolution is predicted to improve by an order of magnitude. The simulated relative energy resolution of the spectrometer with such a pre-collimating lens is 0.021% for an entrance aperture angular spread of ±6°, and 0.088% for an entrance angular spread of ±10°.
Keywords :
Electron energy spectrometer , Scanning electron microscopy , Toroidal spectrometer
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2011
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2208531