• Title of article

    Parabolic versus elliptic focusing – Optimization of the focusing design of a cold triple-axis neutron spectrometer by Monte-Carlo simulations

  • Author/Authors

    Komarek، نويسنده , , A.C. and Bِni، نويسنده , , P. and Braden، نويسنده , , M.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    10
  • From page
    63
  • To page
    72
  • Abstract
    We present Monte-Carlo simulations for the focusing design of a novel cold-neutron triple-axis spectrometer to be installed at the end position of the cold guide NL-1 of the research reactor FRM-II in Munich, Germany. Our simulations are of general relevance for the design of triple-axis spectrometers at end positions of neutron guides. Using the McStas program code we performed ray trajectories to compare parabolic and elliptic focusing concepts. In addition the design of the monochromator was optimized concerning crystal size and mosaic spread. The parabolic focusing concept is superior to the elliptic alternative in view of the neutron intensity distribution as a function of energy and divergence. In particular, the elliptical configuration leads to an inhomogeneous divergence distribution.
  • Keywords
    Elliptic , Neutron guide , focusing , McStas , Supermirror , triple-axis spectrometer , parabolic
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2011
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number