The measurement of the presampled MTF of a high spatial resolution neutron imaging system
Author/Authors
Cao، نويسنده , , Raymond Lei. and Biegalski، نويسنده , , Steven R.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
8
From page
621
To page
628
Abstract
A high spatial resolution neutron imaging device was developed at the Mark II TRIGA reactor at The University of Texas at Austin. As the modulation transfer function (MTF) is recognized as a well-established parameter for evaluation of imaging system resolution, the aliasing associated with digital sampling adds complexity to its measurement. Aliasing is especially problematic when using a high spatial resolution micro-channel plate (MCP) neutron detector that has a pixel grid size similar to that of a CCD array. To compensate for the aliasing an angulated edge method was used to evaluate the neutron imaging facility, overcoming aliasing by obtaining an oversampled edge spread function (ESF). Baseline correction was applied to the ESF to remove the noticeable trends and the LSF was multiplied by Hann window to obtain a smoothed version of presampled MTF. The computing procedure is confirmed by visual inspection of a testing phantom; in addition, it is confirmed by comparison to the MTF measurement of a scintillation screen with a known MTF curve.
Keywords
Spatial resolution , Presampled MTF , Edge method , MCP neutron detector , Neutron imaging