Title of article :
Measurement of decay time of liquid scintillator
Author/Authors :
Zhong، نويسنده , , Wei-Li and Li، نويسنده , , Zu-Hao and Yang، نويسنده , , Chang-Gen and Cao، نويسنده , , Jun، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Abstract :
In this paper we describe a single photon method to measure the decay time of two kinds of homemade liquid scintillator. We find that the fast and slow decay time constants of these two samples, with 5 cm diameter and 4 cm height, are about ∼ 4 ns and ∼ 17 ns , respectively. These results are checked through a reference measurement by recording the scintillation pulse shapes at the output of a fast response PMT directly by a digital oscilloscope. The time responses of these two samples are as good as those of commercial liquid scintillator.
Keywords :
Single photon method , Reactor neutrino oscillation , Liquid scintillator , Decay time
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A