Title of article :
Evaluation of lead shielding for a gamma-spectroscopy system
Author/Authors :
Smith، نويسنده , , M.L. and Bignell، نويسنده , , L. and Alexiev، نويسنده , , D. and Mo، نويسنده , , L. and Harrison، نويسنده , , J.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Abstract :
Three types of lead shielding assemblies were investigated to establish which provides the lowest background as preliminary work for determining the efficiency of a High Purity Germanium (HPGe) detector. The background continuum and peaks were examined using gamma spectroscopy with a coaxial and a planar HPGe detector. The 210Pb concentration in each lead sample taken from the assemblies was quantitatively determined. A French lead sample was found to have the lowest background in this relative comparison with a 210Pb concentration of 29.1±1.3 Bq kg−1.
Keywords :
Lead Shield , Alpha spectrometry , Gamma Spectrometry , 210Pb
Journal title :
Nuclear Instruments and Methods in Physics Research Section A