Title of article :
An axis of rotation alignment system for high-resolution pinhole SPECT imaging
Author/Authors :
Kim، نويسنده , , Joong-Hyun and Lee، نويسنده , , Jae-Sung and Park، نويسنده , , Min-Jae and Lee، نويسنده , , Won Woo and Kim، نويسنده , , Soo Mee and Kim، نويسنده , , Yu Kyeong and Kim، نويسنده , , Sang-Eun and Lee، نويسنده , , Dong Soo، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
7
From page :
338
To page :
344
Abstract :
We developed a simple geometric calibration method for modified animal pinhole single photon emission computed tomography (SPECT) imaging systems using a laser alignment system that does not require additional calibration scans, and verified the feasibility of this system. An optical system consisting of a laser generator and a reflecting mirror, and an alignment method using this system were developed. After laser alignment had been completed, SPECT scans of a Tc-99m line source were performed. After acquiring data with complete alignment, the rotation stage was moved along the y-axis (parallel to the detector plane) to acquire off-axis data. Using these on- and off-axis settings, a micro-performance phantom with hot-rod inserts was scanned after filling with Tc-99m solution. Reconstructed images of the line source and the hot-rod phantom showed that image degradation was minimized when the rotation center was completely aligned using our system. The spatial resolution of the reconstructed image measured using the line source was finest under complete alignment. Under this condition, the smallest hot rod (1.2 mm diameter) was resolved in the SPECT image acquired using a 0.5 mm pinhole aperture. The effects of misalignment were clearly observable in off-axis images; only hot rods with more than 3.2 mm diameter were resolved in 1 mm off-axis image whereas there were no resolvable hot rods with more than 2 mm off-axis image. This system of which the feasibility was verified in the present study will be useful for low-cost molecular imaging studies using single photon emitting tracers.
Keywords :
SPECT , molecular imaging , Pinhole collimator , Center of rotation error
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2008
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2209529
Link To Document :
بازگشت