Title of article :
Digital pulse processing for planar TlBr detectors
Author/Authors :
Nakhostin، نويسنده , , M. and Hitomi، نويسنده , , K. and Ishii، نويسنده , , K. and Kikuchi، نويسنده , , Y.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Abstract :
We report on a digital pulse processing algorithm for correction of charge trapping in the planar TlBr detectors. The algorithm is performed on the signals digitized at the preamplifier stage. The algorithm is very simple and is implemented with little computational effort. By using a digitizer with a sampling rate of 250 MSample/s and 8 bit resolution, an energy resolution of 6.5% is achieved at 511 keV with a 0.7 mm thick detector.
Keywords :
digital signal processing , ?-ray spectrometers , Thallium bromide
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A