Title of article :
Increasing recoil mass spectrometers’ resolution by using crossed-field analyzers
Author/Authors :
Ioanoviciu، نويسنده , , Damaschin، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
5
From page :
490
To page :
494
Abstract :
The mass dispersion of the QQEBEQQ symmetric recoil mass spectrometers can be improved substantially, so potentially their resolution, by inserting cylindrical plates inside the magnet gap, thus creating there a crossed field analyzer. The conditions to keep stigmatic and energy focusing were derived for such a modified system. Applied to some specific cases these formulae predict an increase of even 200% in mass dispersion for an increase of total charged particle path by 7.92%. The mass separation of equal momentum beams at middle of the instrument and the resolution-beam acceptance product were calculated for two representative separator configurations.
Keywords :
Crossed-field analyzer , Mass dispersion , Recoil Mass Spectrometer , Recoil mass separator , Ion Optics
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2009
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2209839
Link To Document :
بازگشت