Title of article
Evaluation and correction of readout artifacts from flat panel detectors for non-destructive testing purposes
Author/Authors
Burtzlaff، نويسنده , , S. and Voland، نويسنده , , V. and Salamon، نويسنده , , M. and Hofmann، نويسنده , , Th. and Uhlmann، نويسنده , , N.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2009
Pages
4
From page
183
To page
186
Abstract
Flat panel detectors are commonly used for non-destructive testing purposes using X-ray technology. During a series of measurements with high absorbing objects, we observed an unknown kind of artifacts especially apparent with high contrast edges. These artifacts lead to unwanted results in radioscopic and computed tomography inspection. Given the object is fully occupying the lower part and half of the upper part of the detector. Looking at the image with high contrast visualization, it can be seen that the covered upper part of the detector is brighter than the covered lower half. The horizontal border of the detector tile is clearly recognizable. Furthermore, the uncovered area directly above the object is darker than next to the edge. In this area the vertical border of the edge below can be localized. We examined and evaluated the effect and developed a correction algorithm. The effect and its correction results are presented.
Keywords
image quality , Non-destructive testing , Artifacts , Flat panel detectors
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2009
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2210008