• Title of article

    Radiogram enhancement and linearization using the beam hardening correction method

  • Author/Authors

    Vavrik، نويسنده , , Daniel and Jakubek، نويسنده , , Jan، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    3
  • From page
    212
  • To page
    214
  • Abstract
    Traditionally, radiography is burdened by the fact that standard X-ray tube spectra are polychromatic and attenuation process in matter is energy dependent, where low-energy X-rays are preferentially absorbed. This beam hardening effect has several consequences: radiograms look noisy especially when low-energy photons are detected, empty holes look filled in computed tomography and observed object looks flatter than it should be. It will be shown it this paper that these problems can be solved using the “linearized signal to equivalent thickness” beam hardening correction method.
  • Keywords
    Beam hardening correction , Flat-field correction , digital radiography