Abstract :
An optimized readout configuration based on a charge preamplifier with pulsed-reset has been designed for Silicon Drift Detectors (SDDs) to be used in Particle Induced X-ray Emission (PIXE) measurements. The customized readout electronics is able to manage the large pulses originated by the protons backscattered from the target material that would otherwise cause significant degradation of X-ray spectra and marked increase in dead time. In this way, the excellent performance of SDDs can be exploited in high-quality proton-induced spectroscopy of low- and medium-energy X-rays. This paper describes the designed readout architecture and the performance characterization carried out in a PIXE setup with MeV proton beams.
Keywords :
Charge preamplifier , PIXE , Silicon drift detectors , Front-end electronics , Backscattered protons