Title of article :
High-Resolution Imaging System (HiRIS) based on H9500 PSPMT
Author/Authors :
Trotta، نويسنده , , C. and Massari، نويسنده , , R. and Trinci، نويسنده , , G. and Palermo، نويسنده , , N. and Boccalini، نويسنده , , S. and Scopinaro، نويسنده , , F. and Soluri، نويسنده , , A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Abstract :
The H8500 PhotoMultiplier Tube (PMT) from Hamamatsu has been used in the last years to assemble several scintigraphic devices in order to achieve high-resolution gamma cameras. If the detector is coupled to discrete scintillator with millimetric pixel size, the resulting charge distribution that emerges is not properly sampled by its anodes (6×6 mm2).
w position sensitive PMT H9500, with its 3×3 mm2 anodes, allows a better charge distribution sampling, improving both spatial resolution and linearity of the system.
s paper, we investigate the imaging performances of the H9500 PMT coupled with a CsI(Tl) array having 1 mm pixel size and compare the results with the same scintillator coupled with H8500 PMT. A portable imaging system named HiRIS (High-Resolution Imaging System) was then realized using a miniaturized readout electronic. Thanks to its lightness, it can be easily used in Medical Imaging. We used HiRIS, together with a rotating system, to carry out a tomographic reconstruction of the biodistribution of a radiopharmaceutical in rats.
Keywords :
nuclear medicine imaging , scintillation detector , Position sensitive photomultiplier tube
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A