Title of article :
Thick Si(Li) coaxial detectors for registration of intermediate energy heavy ions
Author/Authors :
Popeko، نويسنده , , L.A. and Kotina، نويسنده , , I.M. and Shishkina، نويسنده , , G.A. and Grigorieva، نويسنده , , L.A. and Artyukh، نويسنده , , A.G. and Teterev، نويسنده , , Yu.G. and Sereda، نويسنده , , Yu.M.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
3
From page :
235
To page :
237
Abstract :
Thick coaxial Si(Li) detectors are fabricated to register long-range high-ionizing reaction products with high energy resolution. It is shown that charge collection efficiency and energy resolution strongly depend on cooling temperature and magnitude of voltage applied. The test experiments have shown that coaxial detectors are able to register the intermediate energy heavy ions with 0.5–0.7% resolution. The energy losses in entrance window of the detector are <100 keV.