Title of article :
Drift chamber for the CMD-3 detector
Author/Authors :
Grancagnolo، نويسنده , , F. Di Fiore، نويسنده , , G. and Ignatov، نويسنده , , F.V. and Karavdina، نويسنده , , A.V. and Khazin، نويسنده , , B.I. and Miccoli، نويسنده , , A. and Pivovarov، نويسنده , , S.G. and Popov، نويسنده , , A.S. and Ruban، نويسنده , , A.A. and Sibidanov، نويسنده , , A.L. and Snopkov، نويسنده , , I.G.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
2
From page :
105
To page :
106
Abstract :
The construction, digitizing electronics and results of test with cosmic rays of cylindrical drift chamber for CMD-3 (Cryogenic Magnetic Detector) detector are described. To uniformly fill sensitive volume small hexagonal cells were chosen. A coordinate along wire is measured by charge division technique. Resolution in R – φ -plane better than 100 μ m and about 2 mm along wire as well as d E / d x resolution about 12% in cosmic test runs were obtained.
Keywords :
Cosmic ray test , CMD-3 , Drift chamber
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2009
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2210535
Link To Document :
بازگشت