Title of article :
Detector for coherent synchrotron radiation measurements from separate electron bunches in a millimeter wavelength region
Author/Authors :
Naumenko، نويسنده , , G. and Potylitsyn، نويسنده , , A. and Kube، نويسنده , , G. and Grimm، نويسنده , , David O. and Cha، نويسنده , , V. and Popov، نويسنده , , Yu.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Abstract :
Room-temperature detector, manufactured by the company “SPINOR” (Tomsk Russia) was tested for the investigations of the coherent synchrotron radiation characteristics at the bent electron beam of FLASH (Hamburg) in the wavelength region 0.5–3 mm. Detector is based on a low-threshold diode and a broadband antenna with a built-in preamplifier. The detector characteristics allowed us to measure a radiation pulse from separate electron bunches from electron train. The measured coherent synchrotron radiation spectral characteristics in comparison with ones, measured using the cryogenic bolometer show a good homogeneity of the spectral sensitivity of our detector for wavelengths longer than 1.5 mm. The comparison of temporal characteristics with the well-established pyro-electric detectors shows a response time around 1.5 ns. The measured detector sensitivity was about 100 MV/J.
Keywords :
detector , Millimeter wavelength , Synchrotron radiation
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A