• Title of article

    Wavelength-shifting-fibre-based neutron image detector with a fibre-optic taper to increase the spatial resolution

  • Author/Authors

    Nakamura، نويسنده , , T. and Katagiri، نويسنده , , M. H. Toh، نويسنده , , K. and Sakasai، نويسنده , , K. and Ebine، نويسنده , , M. and Birumachi، نويسنده , , A. and Soyama، نويسنده , , K.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    3
  • From page
    158
  • To page
    160
  • Abstract
    The effective pixel size of a two-dimensional wavelength-shifting-fibre (WLS-fibre)-based neutron image detector was improved from 0.5 to 0.16 mm by including a fibre-optic taper (FOT) between the scintillator screen and the WLS fibre. The WLS-fibre-based detector consisted of a thin ZnS/6LiF screen, a FOT and WLS ribbons crossed in the x and y directions. The demonstrator detector had 16 fibre channels in each direction, and the light signals in each fibre were read out individually. The FOT was constructed from fine glass fibres with a taper ratio of 3.1 and served as an image magnifier. The prototype detector equipped with the FOT exhibited a spatial resolution of 0.3 mm, compared to the spatial resolution of 0.8 mm of the original detector without the FOT.
  • Keywords
    neutron detector , Wavelength-shifting-fibre , Fibre-optic taper , High spatial resolution
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2009
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number