Title of article :
Reborn quadrant anode image sensor
Author/Authors :
Prokazov، نويسنده , , Yury and Turbin، نويسنده , , Evgeny and Vitali، نويسنده , , Marco and Herzog، نويسنده , , Andreas and Michaelis، نويسنده , , Bernd and Zuschratter، نويسنده , , Werner and Kemnitz، نويسنده , , Klaus، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
3
From page :
221
To page :
223
Abstract :
We describe a position sensitive photon counting microchannel plate based detector with an improved quadrant anode (QA) readout system. The technique relies on a combination of the four planar elements pattern and an additional fifth electrode. The charge cloud induced by single particle detection is split between the electrodes. The measured charge values uniquely define the position of the initial event. QA has been first published in 1976 by Lampton and Malina. This anode configuration was undeservedly forgotten and its potential has been hardly underestimated. The presented approach extends the operating spatial range to the whole sensitive area of the microchannel plate surface and demonstrates good linearity over the field of view. Therefore, the novel image sensor results in spatial resolution better then 50 μ m and count rates up to one million events per second.
Keywords :
Imaging , Quadrant anode , Image distortion corrections , Microchannel-plate , Position sensitive photomultiplier
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2009
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2211221
Link To Document :
بازگشت