• Title of article

    Scintillation fiber array detector for measurement of neutron beam profile

  • Author/Authors

    Kim، نويسنده , , Chong and Hong، نويسنده , , Byungsik and Jo، نويسنده , , Mihee and Lee، نويسنده , , Kyong Sei and Sim، نويسنده , , Kwang-Souk، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    5
  • From page
    276
  • To page
    280
  • Abstract
    We built and tested a detector to measure the profile of fast-neutron beams delivered by the MC50 cyclotron at the Korea Institute of Radiological and Medical Science (KIRAMS). The core component of the detector is a 2 × 46 array of scintillation fibers. The light output of the scintillation fibers is transformed into a current signal by a 46-channel silicon photodiode and digitized by a current-mode signal processor. This scanning device was designed to cover a neutron beam area of 30 × 32 cm 2 . The detector was tested in a neutron beam delivered by the MC50 cyclotron at KIRAMS. We demonstrate that the detector can successfully measure the neutron beam profile at various beam currents from 10 to 20 μ A . The proposed neutron beam profile detector will be useful, for example, in radiotherapy applications with neutron intensities above 10 7 Hz / cm 2 .
  • Keywords
    Line scanning , Neutron beam , Scintillation fiber , Current-mode electronics , Proton accelerator
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2009
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number