Title of article
Displacement Uncertainty in Interferometric Radius Measurements
Author/Authors
Schmitz، نويسنده , , T.L. and Evans، نويسنده , , C.J. and Davies، نويسنده , , A. and Estler، نويسنده , , W.T.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2002
Pages
4
From page
451
To page
454
Abstract
Interferometric radius measurements may be completed using a radius bench, where radius is defined as the displacement between the confocal and catʹs eye nulls (identified using a figure measuring interferometer). Measurements of a Zerodur sphere have been completed on the X-ray Optics Calibration interferometer (XCALIBIR) and a coordinate measuring machine. Larger recorded disagreements than indicated by the current uncertainty analysis call for an exploration of the analysis model. This paper details uncertainties associated with the use of multiple displacement measuring interferometers (DMIs) to record motion in a single axis by treating the specific case of displacement measurement on XCALIBIR using three DMIs equally spaced around the optical axis.
Keywords
Radius measurement , INTERFEROMETRY , uncertainty
Journal title
CIRP Annals - Manufacturing Technology
Serial Year
2002
Journal title
CIRP Annals - Manufacturing Technology
Record number
2266437
Link To Document