Title of article :
Rapid measurement of calcite concentration in copper ore using fluorescence spectroscopy
Author/Authors :
Escلrate، نويسنده , , P. and Guesalaga، نويسنده , , A. and Otero، نويسنده , , A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
4
From page :
1373
To page :
1376
Abstract :
This work presents the development of a rapid technique for measuring the concentration of calcite in oxidized copper ore. The instrument is based on fluorescence spectroscopy in the visible and infrared range and is excited with a UV-LW lamp. The measurement takes substantially shorter times than standard procedures. ation models were adjusted via the correlation coefficient for validation ( R v 2 ) and the error of prediction (RMSEP). Correlations higher than 0.92 were achieved for R v 2 . These results show the potential of the technique for portable or on-line estimation of calcite concentration in mining processes. Its performance under different granulometries is also addressed, showing the best performance for particle size between 0.01 and 0.04 mm.
Keywords :
Copper , Granulometry , calcite , Spectroscopy , fluorescence
Journal title :
Minerals Engineering
Serial Year :
2006
Journal title :
Minerals Engineering
Record number :
2274726
Link To Document :
بازگشت