Author/Authors :
Sajjadieh-Kajouei، Amirreza نويسنده Cardiologist, Cardiac Rehabilitation Research Center, Isfahan Cardiovascular Research Institute, Isfahan University of Medical Sciences, Isfahan, Iran , , Hekmatnia، Ali نويسنده , , Keivani، Maryam نويسنده Resident, Shahrekord University of Medical Sciences, Shahrekord, Iran , , Asoodeh، Abdollah نويسنده Resident, Cardiac Rehabilitation Research Center, Isfahan Cardiovascular Research Institute, Isfahan University of Medical Sciences, Isfahan, Iran , , Pourmoghaddas، Masoud نويسنده MD, Professor of Interventional Cardiology, Chamran Heart Hospital, Isfahan University of Medical Sciences, Isfahan , , Sanei، Hamid نويسنده Department of Internal Medicine ,
Abstract :
BACKGROUND: The aim of the present study is to evaluate the accuracy of 64-multidetector-row
computed tomography angiography (CTA) in comparison to conventional invasive angiography
(CIA) in the diagnosis of significant stenosis (? 50%) of coronary artery tree.
METHODS: Assessment of CTA in the detection of coronary artery disease (CAD) was performed
in patients referred because of symptoms or stress studies suggestive of ischemia. For this
purpose, among more than 1000 cases of coronary CTA in a 20 months period a study
population of 54 patients suspected to have significant stenosis of the coronary artery tree was
investigated. The CIA procedure was performed in these patients one month after CTA. The
accuracy of CTA in detecting significant stenosis was compared to CIA.
RESULTS: For vessel based analysis of 179 coronary vessels, CTA had a sensitivity of 96%,
specificity of 87.5%, positive predictive value of 90.5%, and negative predictive value of 94.6%.
For patient-base analysis, the sensitivity, specificity, positive predictive value (PPV), and
negative predictive value (NPV) of CTA were 97.9%, 28.6%, 66.6%, and 90.2%, respectively.
CONCLUSION: The findings of this study reveal that CT angiography with 64-slice scanner could
be considered as a suitable technique for rapid triage of patients presenting to hospitals with chest
pain. High values of sensitivity and PPV reveal the good performance of CTA in detecting CAD.