Title of article
Failure of an epidural catheter
Author/Authors
Lewis، نويسنده , , Peter R. and Gagg، نويسنده , , Colin، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2009
Pages
11
From page
1805
To page
1815
Abstract
Good product design implies integrity when a product is in service, and is especially important in medical devices. The environment in which the product functions can vary but must be taken into account as well as its past history of environmental exposure. A case study shows how an epidural catheter fractured during childbirth, leaving the distal tip in the spinal fluid of the patient. Analysis of the proximal remains suggested that the polymer had become brittle. FTIR microscopy showed chain oxidation with UV degradation as the most likely cause of the problem. An allegation that the anaesthetist had been negligent was shown to be groundless. Many common polymers are susceptible to UV cracking, and precautions are needed during manufacture and assembly to prevent failure when the device is needed to perform a critical job.
Keywords
Catheter , Polymer , Degradation , Failure
Journal title
Engineering Failure Analysis
Serial Year
2009
Journal title
Engineering Failure Analysis
Record number
2338592
Link To Document