Title of article :
Electron cyclotron emission measurements by means of a grating polychromator on the large helical device
Author/Authors :
de Vries، نويسنده , , P.C and Kawahata، نويسنده , , K and Nagayama، نويسنده , , Y and Inagaki، نويسنده , , S and Sasao، نويسنده , , H and Ito، نويسنده , , Y، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
7
From page :
185
To page :
191
Abstract :
The electron cyclotron emission (ECE) spectrum at the large helical device (LHD) is measured by a 14-channel grating polychromator. During standard operation, the polychromator monitors second harmonic frequencies (100–150 GHz). At sufficient high density, the second harmonic X-mode polarisation is optically thick and can be used to determine the temperature profile. However, the large magnetic field shear in LHD affects the ECE polarisation. This effect has been studied numerically. The wave polarisation was found to rotate in the laboratory frame. Eperiments have been carried out by means of a polarisation rotator in the diagnostic waveguide system, which confirmed the calculations. By a proper setting of the polarisation rotator, the rotation can be corrected and pure X-mode is detected. Temperature profiles have been measured successfully by the polychromator.
Keywords :
ECE diagnostics , Large helical device , Mode conversion