• Title of article

    Plasma-ion diagnostics using a newly designed differential-spectrum ion-energy analyzer with electrostatic slanted grids

  • Author/Authors

    Sakarnoto، نويسنده , , Y and Cho، نويسنده , , T and Hirata، نويسنده , , M and Yoshimura، نويسنده , , T and Makino، نويسنده , , K and Nakashima، نويسنده , , Y and Yatsu، نويسنده , , K and Tamano، نويسنده , , T and Miyoshi، نويسنده , , S، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    4
  • From page
    543
  • To page
    546
  • Abstract
    Ion energies, temperatures, and ion-confining potentials in open-ended plasma-confinement devices are clearly measured with our newly designed ion-energy spectrometer. This energy spectrometer has specific structures with obliquely positioned multiple grids. Such a grid geometry makes it possible to separate ions and electrons completely. This electrostatic spectrometer also has the following characteristics: (1) no use of internal magnetic fields (being different from E|B spectrometers); and (2) the use of a simple single-channel collector plate for obtaining differential ion-energy spectra (being different from energy-integrated spectra using Faraday cups). These characteristics provide us with easy handling of the spectrometer including an easy alignment of the radially scannable spectrometer in a plasma-confinement device.
  • Journal title
    Fusion Engineering and Design
  • Serial Year
    1997
  • Journal title
    Fusion Engineering and Design
  • Record number

    2363585