• Title of article

    Calibration of the CHS HIBP ion beam trajectory with a gas ionization method

  • Author/Authors

    Lee، نويسنده , , S and Fujisawa، نويسنده , , A and Crowley، نويسنده , , T.P and Iguchi، نويسنده , , H and Kojima، نويسنده , , M and Hamada، نويسنده , , Y and Nishimura، نويسنده , , K and Okamura، نويسنده , , S and Matsuoka، نويسنده , , K، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    4
  • From page
    645
  • To page
    648
  • Abstract
    A three-dimensional complex trajectory is imposed on the heavy ion beam probe (HIBP) particle orbits in heliotron/torsatron devices. A pair of two-dimensional beam deflectors for primary and secondary beam lines are equipped in the Compact Helical System (CHS) HIBP, which makes it possible to measure the plasma potential at almost all plasma radii. A plasma observation point is determined by a set of four deflector voltages. Experimental verification for these observation points is necessary in such a complex configuration. In this study, a movable detector and a gas ionization method have been used to find these deflector voltages. Calibration procedures for these methods are described in detail. The movable detector is also used to optimize the focusing condition of the primary beam in the plasma region.
  • Journal title
    Fusion Engineering and Design
  • Serial Year
    1997
  • Journal title
    Fusion Engineering and Design
  • Record number

    2363704