Title of article :
New device and method for measuring thermal conductivity of thin-films
Author/Authors :
Subramanian، نويسنده , , Chelakara S. and Amer، نويسنده , , Tahani and UpChurch، نويسنده , , Billy T. and Alderfer، نويسنده , , David W. and Burkett، نويسنده , , Cecil and Sealey، نويسنده , , Bradley، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
6
From page :
313
To page :
318
Abstract :
Thermal sensitive paints (TSPs) are used for global nonintrusive detection of boundary layer transition in flow over the surface of wind tunnel research models. Since the transition is a transient process, the TSP should have a fast response characteristic. A low paint thermal conductivity is required for fast response. A thin-film thermal conductivity meter (TFTCM) was designed and built to measure thermal conductivity of the TSPs, which are typically between 50 and 150/μm thick. In this paper, the design and operating features of the TFTCM are described. Measurement of the thermal conductivity with this TFTCM of three standard thin-film low conductivity specimens, Kapton, Teflon, and Borofloat glass, showed good agreement with the manufacturer quoted values, thus validating the instrument and the procedure. Consistently repeatable values for thermal conductivity (k=0.41±0.02 W/mK) were also obtained for the TSP specimen (TSB-B, 75 /μm) tested.
Keywords :
thin-film , TRANSITION , Temperature sensitive paint , Thermal conductivity meter
Journal title :
ISA TRANSACTIONS
Serial Year :
2006
Journal title :
ISA TRANSACTIONS
Record number :
2382754
Link To Document :
بازگشت