Author/Authors :
Eslami، J نويسنده Lecturer of Anesthesiology, Anesthesiology Department, School of Nursing and Midwifery, Shiraz University of Medical Sciences, Shiraz, Iran , , Ghafaripour، F نويسنده Medical Physics & Medical Engineering Department, School of medicine, Shiraz University of Medical Sciences, Shiraz, Iran , , Mortazavi، S A R نويسنده Medical Student, School of medicine, Shiraz University of Medical Sciences, Shiraz, Iran , , Mortazavi، S M J نويسنده Medical Physics & Medical Engineering Department, School of medicine, Shiraz University of Medical Sciences, Shiraz, Iran , , Shojaei-fard ، M B نويسنده Ionizing and Non-ionizing Radiation Protection Research Center, Shiraz University of Medical Sciences, Shiraz, Iran ,
Abstract :
Background: People who use home blood glucose monitors may use their mobile
phones in the close vicinity of medical devices. This study is aimed at investigating
the effect of the signal strength of 900 MHz GSM mobile phones on the accuracy of
home blood glucose monitors.
Methods: Sixty non-diabetic volunteer individuals aged 21 - 28 years participated
in this study. Blood samples were analyzed for glucose level by using a common blood
glucose monitoring system. Each blood sample was analyzed twice, within ten minutes in presence and absence of electromagnetic felds generated by a common GSM
mobile phone during ringing. Blood samples were divided into 3 groups of 20 samples
each. Group 1: exposure to mobile phone radiation with weak signal strength. Group2:
exposure to mobile phone radiation with strong signal strength. Group3: exposure to a
switched–on mobile phone with no signal strength.
Results: The magnitude of the changes in the frst, second and third group between glucose levels of two measurements (??C )?were 7.4±3.9 mg/dl, 10.2±4.5 mg/
dl, 8.7±8.4 mg/dl respectively. The difference in the magnitude of the changes between
the 1st and the 3rd groups was not statistically signifcant. Furthermore, the difference
in the magnitude of the changes between the 2nd and the 3rd groups was not statistically signifcant.
Conclusion: Findings of this study showed that the signal strength of 900 MHz
GSM mobile phones cannot play a signifcant role in changing the accuracy of home
blood glucose monitors.