Title of article :
Investigation of Collimator Influential Parameter on SPECT Image Quality: a Monte Carlo Study
Author/Authors :
Banari Bahnamiri، Sh نويسنده Assistant Professor of Nuclear Physics, Tabari University of Babol, Iran ,
Issue Information :
فصلنامه با شماره پیاپی 0 سال 2015
Pages :
10
From page :
39
To page :
48
Abstract :
Background: Obtaining high quality images in Single Photon Emission Tomography (SPECT) device is the most important goal in nuclear medicine. Because if image quality is low, the possibility of making a mistake in diagnosing and treating the patient will rise. Studying effective factors in spatial resolution of imaging systems is thus deemed to be vital. One of the most important factors in SPECT imaging in nuclear medicine is the use of an appropriate collimator for a certain radiopharmaceutical feature in order to create the best image as it can be effective in the quantity of Full Width at Half Maximum (FWHM) which is the main parameter in spatial resolution. Method: In this research, the simulation of the detector and collimator of SPECT imaging device, Model HD3 made by Philips Co. and the investigation of important factors on the collimator were carried out using MCNP-4c code. Results: The results of the experimental measurments and simulation calculations revealed a relative difference of less than 5% leading to the confrmation of the accuracy of conducted simulation MCNP code calculation. Conclusion: This is the frst essential step in the design and modelling of new collimators used for creating high quality images in nuclear medicine.
Journal title :
Journal of Biomedical Physics and Engineering
Serial Year :
2015
Journal title :
Journal of Biomedical Physics and Engineering
Record number :
2393138
Link To Document :
بازگشت